Alicat MWB Series Low Pressure Drop Battery Meters
Performance
Available ranges: max 0-1000 slpm, min 0-0.5 sccm/0-500 sμlm (see Part Numbers tab for stock ranges)
Operating range (Turndown ratio): 200:1
Standard accuracy calibration, NIST-traceable: +/- (0.8% of the reading + 0.2% of the full-scale range)
Optional high-accuracy calibration, NIST-traceable: +/- (0.4% of the reading + 0.2% of the full-scale range)
Repeatability: 0.2% of the full-scale range
Typical measurement response time constant: 5-10 ms
No warm-up time required
Operating Conditions
Compatible gases: 98 pre-loaded selectable gas calibrations; additional mixtures definable with COMPOSER
Operating temperature: -10°C to +60°C
Max internal pressure (static): 45 psig (60 psia)
Max differential pressure (inlet to outlet): 15 psid
Proof pressure: 175 psig (190 psia)
STP/NTP: user-selectable (defaults to 25°C & 1 atm STP/0°C & 1 atm NTP)