OmniaTap xs basic 5 Ultrapure Water System
Feedwater requirements
Source |
Tap water according to DIN 2000 |
Feedwater temperature |
+2°C to 35° |
Feedwater pressure |
0,5 up to 6 bar |
Conductivity at 25° |
< 2000 μS/cm |
Colloid index SDI |
< 10 |
Free chlorine |
< 3 ppm |
TOC-Value |
< 1 ppm |
Hardness [as CaCO3]* |
< 300 μS/cm |
Ph range |
4 up to 10 |
*For higher values, pre-treatment must be carried out upstream
Type I ultrapure water **
Ultrapure water conductivity |
18.2 MΩ x cm ≙ 0.055 μS/cm |
Dispensing performance |
up to 2 l/min. |
TOC value |
< 10 ppb* |
Particle content |
< 1/ml** |
Bacteria content |
< 0,01 CFU/ml** |
Typical applications |
AAS, IC, ICP, buffers and media preparation |
* Dependent on the feedwater and regular disinfection
**with 0,2 μm Sterile filter capsule 19100300 or Bio filter capsule 19102100
Type II pure water (Tank outlet)
Pure water conductivity |
15-10 MΩ x cm ≙ 0.067-0.1 μS/cm |
Pure water performance at 15°C |
5 l/h |
Typical applications
|
Rinsing of laboratory glass Preparation of reagents and samples Feed water for autoclaves Make up water for buffers and media |
Technical data
Ambient temperature |
+2 to 35°C |
Supply voltage |
90-240 Volt / 50-60 Hz |
Total connected load |
120 W |
Port size – tap water |
R ¾“ - female |
Dimensions |
W 253 x H 530 x D 520 mm |
Weight |
19 kg |
For more detailed information please contact with us. satis@analitika.com.tr