OmniaTap xs basic 5 UV Ultrapure Water System

Download

OmniaTap xs basic 5 UV Ultrapure Water System

Feedwater requirements

Source

Tap water according to DIN 2000

Feedwater temperature

+2°C to 35°

Feedwater pressure

0,5 up to 6 bar

Conductivity at 25°

< 2000 μS/cm

Colloid index SDI

< 10

Free chlorine

< 3 ppm

TOC-Value

< 1 ppm

Hardness [as CaCO3]*

< 300 μS/cm

Ph range

4 up to 10

*For higher values, pre-treatment must be carried out upstream

 

Type I ultrapure water **

Ultrapure water conductivity

18.2 MΩ x cm ≙ 0.055 μS/cm

Dispensing performance

up to 2 l/min.

TOC value

< 5 ppb*

Particle content

< 1/ml**

Bacteria content

< 0,01 CFU/ml**

Typical applications

Ultra-trace analysis, ICP-MS

HPLC, TOC-analysis

* Dependent on the feedwater and regular disinfection

**with 0,2 μm Sterile filter capsule 19100300 or Bio filter capsule 19102100

 

Type II pure water (Tank outlet)

Pure water conductivity

15-10 MΩ x cm ≙ 0.067-0.1 μS/cm

Pure water performance at 15°C

5 l/h

Typical applications

 

Rinsing of laboratory glass

Preparation of reagents and samples

Feed water for autoclaves

Make up water for buffers and media

 

Technical data

Ambient temperature

+2 to 35°C

Supply voltage

90-240 Volt / 50-60 Hz

Total connected load

0.1 kW

Port size – tap water

R ¾“ - female

Dimensions

W 253 x H 530 x D 520 mm

Weight

19 kg

For more detailed information please contact with us. satis@analitika.com.tr