OmniaTap12 Ultrapure Water System
Feedwater requirements
Drinking water according to DIN 2000 |
|
Feedwater temperature |
+2°C to 35° |
Feedwater pressure |
2 to 6 bar |
Manganese and iron content |
< 0.05 mg/l |
Free chlorine content |
< 0,1 mg/l |
Silt density index (SDI) |
max. 3 |
Type I ultrapure water (Hand dispensing)
Ultrapure water conductivity |
18.2 MΩ x cm ≙ 0.055 μS/cm |
Dispensing performance |
up to 2 l/min. |
TOC value |
< 10 ppb* |
Particle content |
< 1 particulate/ml |
Bacteria content |
< 0.01 CFU/ml** |
Typical applications |
AAS, IC, ICP |
*Dependent on the feedwater and regular disinfection
** with sterile filter 0.2 μm
Type II pure water (Tank outlet)
Pure water conductivity |
15-10 MΩ x cm ≙ 0.067-0.1 μS/cm |
Pure water performance at 15°C |
12 l/h |
Typical applications
|
Rinsing of laboratory glass Preparation of reagents and samples Feed water for autoclaves Make up water for buffers and media |
Technical data
Ambient temperature |
+2 to 35°C |
Supply voltage |
90-240 Volt / 50-60 Hz |
Total connected load |
0.1 kW |
Inlet/rinsing hose connector |
d8 mm |
Concentrate connector |
d8 mm |
Dimensions |
W 390 x D 615 x H 720 mm |
Weight |
22 kg |
For more information please contact with us. satis@analitika.com.tr